Fluke ESA 612 with Patient Leakage Tests Electrical Safety Analyzer











>> YOUR LINK HERE: ___ http://youtube.com/watch?v=MA0fkBKDOZI

In this vide we cover how to use a Fluke ESA 612 electrical safety analyzer. We demonstrate ground wire resistance testing, touch currents tests, and patient leakage tests. • The test performed are in line with NFPA 99 - AAMI/ANSI 99 testing standards • Ground wires resistance values should be less than 0.5 ohms • Touch current should be run in • No Fault Condition (ground intact) with the DUT in the on and off state less than 100 microamps • Single Fault Condition (ground lifted) with the DUT in the on and off state less than 500 microamps • Patient Leakage and Patient Auxiliary Leakage • No Fault Condition (ground intact) with the DUT in the on and off state less than 100 microamps • Single Fault Condition (ground lifted) with the DUT in the on and off state less than 500 microamps • Hope you enjoy the video make sure to subscribe to learn more! • Some of my favorite Fluke Products! • Voltage Detector - https://amzn.to/3rfapv9 • Fluke Multimeter and Clamp Combo Kit - https://amzn.to/3xddAax • Fluke VT04A Infrared Imager - https://amzn.to/3E866Y1

#############################









Content Report
Youtor.org / YTube video Downloader © 2025

created by www.youtor.org