BW GasAlert Micro Gas Calibration Guide by Ideal Calibrations











>> YOUR LINK HERE: ___ http://youtube.com/watch?v=aTkdm7sq3Qk

This is a video in the How To Calibrate Your Gas Monitor series by Ideal Calibrations. All you need is trusted calibration gas (of course we recommend Ideal Gases brand), a cal cup, a regulator, and tubing. • If you have a problem during your calibration, give us a call or send it in for repair here: • https://idealcalibrations.com/gas-det... • BW GasAlert Micro Gas Calibration Guide by Ideal Calibrations • 1. Verify gas blend and expiration date. • 2. Turn the unit on. • 3. Press and hold the power button and let it shut down but keep holding it until it starts up in calibration mode. • Calibration Gas for the BW GasAlert Micro is: • 25ppm Hydrogen Sulfide • 100ppm Carbon Monoxide • 50% LEL Methane • 18.0% Oxygen • Nitrogen Balance • Buy the calibration gas here: • https://idealcalibrations.com/calibra... • Regulator - 0.5 lpm • Buy the Regulator here: • https://idealcalibrations.com/0-5-lpm... • If you have any gas monitor or calibration gas questions, please feel free to call us at 1-734-956-0539 or to email me [email protected]. We'd be happy to answer any questions you have. • Please see our websites at: • https://www.idealcalibrations.com

#############################